ABSTRACT: Background
We aimed to evaluate the analytic performance of three rapid HIV viral load assays: the novel Xpert HIV-1 VL XC (Xpert XC), the Xpert HIV-1 VL (Xpert VL), and the m-PIMA HIV-1/2 VL (m-PIMA). Setting
Two South African clinics. Methods
We conducted a prospective diagnostic accuracy study. Site-laboratory technicians and nurses used the Xpert XC, Xpert VL and m-PIMA to test plasma samples from people with HIV receiving antiretroviral therapy. We compared results with the Roche Cobas HIV-1 reference assay. We determined accuracy to detect viraemia at the World Health Organization (WHO) failure threshold of 1000 copies/mL on all three assays, and 50 and 200 copies/mL on the Xpert assays. We assessed agreement using Bland-Altman plots. Results
We en