Diagnostic accuracy of the rapid Xpert HIV-1 Viral Load XC, Xpert HIV-1 Viral Load & m-PIMA HIV-1/2 Viral Load in South African clinics


ABSTRACT:

Background

We aimed to evaluate the analytic performance of three rapid HIV viral load assays: the novel Xpert HIV-1 VL XC (Xpert XC), the Xpert HIV-1 VL (Xpert VL), and the m-PIMA HIV-1/2 VL (m-PIMA).

Setting

Two South African clinics.

Methods

We conducted a prospective diagnostic accuracy study. Site-laboratory technicians and nurses used the Xpert XC, Xpert VL and m-PIMA to test plasma samples from people with HIV receiving antiretroviral therapy. We compared results with the Roche Cobas HIV-1 reference assay. We determined accuracy to detect viraemia at the World Health Organization (WHO) failure threshold of 1000 copies/mL on all three assays, and 50 and 200 copies/mL on the Xpert assays. We assessed agreement using Bland-Altman plots.

Results

We en