Whole brain 31 P MRSI at 7T with a dual-tuned receive array.


ABSTRACT:

Purpose

The design and performance of a novel head coil setup for 31 P spectroscopy at ultra-high field strengths (7T) is presented. The described system supports measurements at both the 1 H and 31 P resonance frequencies.

Methods

The novel coil consists of 2, actively detunable, coaxial birdcage coils to give homogeneous transmit, combined with a double resonant 30 channel receive array. This allows for anatomical imaging combined with 31 P acquisitions over the whole head, without changing coils or disturbing the subject. A phosphate buffer phantom and 3 healthy volunteers were scanned with a pulse acquire CSI sequence using both the novel array coil and a conventional transceiver birdcage. Four different methods of combining the array channels were compared at 3 differen