Specific absorption rate and temperature in neonate models resulting from exposure to a 7T head coil.


ABSTRACT:

Purpose

To investigate safe limits for neonatal imaging using a 7T head coil, including both specific absorption rate (SAR) and temperature predictions.

Methods

Head-centered neonate models were simulated using finite-difference time domain-based electromagnetic and thermal solvers. The effects of higher water content of neonatal tissues compared with adults, position shifts, and thermal insulation were also considered. An adult model was simulated for comparison.

Results

Maximum and average SAR are both elevated in the neonate when compared with an adult model. When normalized to B1+ , the SAR experienced by a neonate is greater than an adult by approximately a factor of 2; when normalized to net forward power (forward-reflected), this increases to a factor of 2.5-3.