Ontology highlight
ABSTRACT: Objective
The aim of this study was to compare brain perfusion SPECT obtained from a 360° CZT and a conventional Anger camera.Methods
The 360° CZT camera utilizing a brain configuration, with 12 detectors surrounding the head, was compared to a 2-head Anger camera for count sensitivity and image quality on 30-min SPECT recordings from a brain phantom and from 99mTc-HMPAO brain perfusion in 2 groups of 21 patients investigated with the CZT and Anger cameras, respectively. Image reconstruction was adjusted according to image contrast for each camera.Results
The CZT camera provided more than 2-fold increase in count sensitivity, as compared with the Anger camera, as well as (1) lower sharpness indexes, giving evidence of higher spatial resolution, for both p
SUBMITTER: Bordonne M
PROVIDER: S-EPMC7642149 | biostudies-literature | 2020 Nov
REPOSITORIES: biostudies-literature