Per-fraction positional and dosimetric performance of prone breast tangential radiotherapy on Halcyon™ linear accelerator assessed with daily rapid kilo-voltage cone beam computed tomography: a single-institution pilot study.


ABSTRACT:

Background

This study investigates daily breast geometry and delivered dose to prone-positioned patients undergoing tangential whole breast radiation therapy (WBRT) on an O-ring linear accelerator with 6X flattening filter free mode (6X-FFF), planned with electronic compensation (ECOMP) method. Most practices rely on skin marks or daily planar image matching for prone breast WBRT. This system provides low dose daily CBCT, which was used to study daily robustness of delivered dose parameters for prone-positioned WBRT.

Methods

Eight patients treated with 16-fraction prone-breast WBRT were retrospectively studied. Planning CTs were deformed to daily CBCT to generate daily synthetic CTs, on which delivered dose distributions were calculated. A total of 8 × 16 = 128 synthetic CTs