Evaluating the serological status of COVID-19 patients using an indirect immunofluorescent assay, France.
Ontology highlight
ABSTRACT: An indirect in-house immunofluorescent assay was developed in order to assess the serological status of COVID-19 patients in Marseille, France. Performance of IFA was compared to a commercial ELISA IgG kit. We tested 888 RT-qPCR-confirmed COVID-19 patients (1302 serum samples) and 350 controls including 200 sera collected before the pandemic, 64 sera known to be associated with nonspecific serological interference, 36 sera from non-coronavirus pneumonia and 50 sera from patient with other common coronavirus to elicit false-positive serology. Incorporating an inactivated clinical SARS-CoV-2 isolate as the antigen, the specificity of the assay was measured as 100% for IgA titre ≥ 1:200, 98.6% for IgM titre ≥ 1:200 and 96.3% for IgG titre ≥ 1:100 after testing a series of negative controls. I