Validation of the BD FACSPresto system for the measurement of CD4 T-lymphocytes and hemoglobin concentration in HIV-negative and HIV-positive subjects.
Ontology highlight
ABSTRACT: This study aimed to compare the performance of the BD FACSPresto system with the conventional standard-of-care technologies for the measurement of absolute CD4 count (AbsCD4), CD4 percentage (CD4%) and total hemoglobin concentration (Hb) in capillary and venous blood samples of HIV-negative and HIV-positive subjects. A total of 1304 participants were included in this prospective cohort study. Both venous and capillary blood samples were analyzed using the BD FACSPresto system and the results were compared against the BD FACSCalibur for enumerating AbsCD4 and CD4% and Sysmex XT-4000i hematology analyzer for determining Hb levels. Method comparison studies were performed using Deming regression and Bland-Altman plots. The Deming regression analyses comparing the accuracy of the BD FACSPresto
SUBMITTER: Lu X