Unknown

Dataset Information

Temperature Correction to Enhance Blood Glucose Monitoring Accuracy Using Electrical Impedance Spectroscopy.


ABSTRACT: Electrical methods are among the primarily studied non-invasive glucose measurement techniques; however, various factors affect the accuracy of the sensors used. Of these, the temperature is a critical factor; hence, the effects of temperature on the electrical properties of blood components are investigated in this study. Furthermore, the changes in the electrical properties of blood according to the glucose level are corrected by considering the effects of temperature on the electrical properties. An impedance sensor is developed and used to measure whole blood impedance in 10 healthy participants at various temperatures and glucose levels. Subsequently, the conductivities of the plasma and cytoplasm were extracted. Changes in the electrical properties of the blood components are then an

SUBMITTER: Lee YS 

PROVIDER: S-EPMC7663582 | biostudies-literature | 2020 Oct

REPOSITORIES: biostudies-literature