Facial Fabrication and Characterization of Novel Ag/AgCl Chloride Ion Sensor Based on Gel-Type Electrolyte.


ABSTRACT: In this study, a novel chloride ion (Cl-) sensor based on Ag wire coated with an AgCl layer was fabricated using a gel-type internal electrolyte and a diatomite ceramic membrane, which played an important role in preventing electrolyte leakage from the ion-selective electrode. The sensing performance, including reversibility, response, recovery time, low detection limit, and the long-term stability, was systemically investigated in electrolytes with different Cl- contents. The as-fabricated Cl- sensor could detect Cl- from 1 to 500 mM KCl solution with good linearity. The best response and recovery time obtained for the optimized sensor were 0.5 and 0.1 s, respectively, for 10 mM KCl solution. An exposure period of over 60 days was used to evalua

SUBMITTER: Kim S 

PROVIDER: S-EPMC7670060 | biostudies-literature | 2020

REPOSITORIES: biostudies-literature