Ontology highlight
ABSTRACT: Purpose
The purpose of this study is to safely acquire the first human head images at 10.5T.Methods
To ensure safety of subjects, we validated the electromagnetic simulation model of our coil. We obtained quantitative agreement between simulated and experimental B1+ and specific absorption rate (SAR). Using the validated coil model, we calculated radiofrequency power levels to safely image human subjects. We conducted all experiments and imaging sessions in a controlled radiofrequency safety lab and the whole-body 10.5T scanner in the Center for Magnetic Resonance Research.Results
Quantitative agreement between the simulated and experimental results was obtained including S-parameters, B1+ maps, and SAR. We calculated peak 10 g average SAR using 4 different realisti
SUBMITTER: Sadeghi-Tarakameh A
PROVIDER: S-EPMC7695227 | biostudies-literature | 2020 Jul
REPOSITORIES: biostudies-literature