Comparison of global end-diastolic volume index derived from jugular and femoral indicator injection: a prospective observational study in patients equipped with both a PiCCO-2 and an EV-1000-device.
Ontology highlight
ABSTRACT: Transpulmonary thermodilution (TPTD)-derived global end-diastolic volume index (GEDVI) is a static marker of preload which better predicted volume responsiveness compared to filling pressures in several studies. GEDVI can be generated with at least two devices: PiCCO and EV-1000. Several studies showed that uncorrected indicator injection into a femoral central venous catheter (CVC) results in a significant overestimation of GEDVI by the PiCCO-device. Therefore, the most recent PiCCO-algorithm corrects for femoral indicator injection. However, there are no systematic data on the impact of femoral indicator injection for the EV-1000 device. Furthermore, the correction algorithm of the PiCCO is poorly validated. Therefore, we prospectively analyzed 14 datasets from 10 patients with TPTD-moni