Foot-Worn Inertial Sensors Are Reliable to Assess Spatiotemporal Gait Parameters in Axial Spondyloarthritis under Single and Dual Task Walking in Axial Spondyloarthritis.
Ontology highlight
ABSTRACT: The aim of this study was (1) to evaluate the relative and absolute reliability of gait parameters during walking in single- and dual-task conditions in patients with axial spondyloarthritis (axSpA), (2) to evaluate the absolute and relative reliability of dual task effects (DTE) parameters, and (3) to determine the number of trials required to ensure reliable gait assessment, in patients with axSpA. Twenty patients with axSpa performed a 10-m walk test in single- and dual-task conditions, three times for each condition. Spatiotemporal, symmetry, and DTE gait parameters were calculated from foot-worn inertial sensors. The relative reliability (intraclass correlation coefficients-ICC) and absolute reliability (standard error of measurement-SEM and minimum detectable change-MDC) were calcula