Multiple MitraClips: The balancing act between pressure gradient and regurgitation.


ABSTRACT:

Objective

Transcatheter mitral valve repair with the MitraClip is used for the symptomatic management of mitral regurgitation (MR). The challenge is reducing MR while avoiding an elevated mitral valve gradient (MVG). This study assesses how multiple MitraClips used to treat MR can affect valve performance.

Methods

Six porcine mitral valves were assessed using an in vitro left heart simulator in the native, moderate-to-severe MR, and severe MR cases. MR cases were tested in the no-MitraClip, 1-MitraClip, and 2-MitraClip configurations. Mitral regurgitant fraction (MRF), MVG, and effective orifice area (EOA) were quantified.

Results

Native MRF, MVG, and EOA were 14.22%, 2.59 mm Hg, and 1.64 cm2, respectively. For moderate-to-severe MR, MRF, MVG, and EOA were

SUBMITTER: Gooden SC 

PROVIDER: S-EPMC7704541 | biostudies-literature | 2022 Apr