ABSTRACT: Background
Double lumen endotracheal tubes (DLT) are commonly used to provide single lung ventilation during thoracic surgery. A fiberoptic bronchoscope (FOB) is typically used to confirm accurate DLT placement. Accounting for initial purchase, maintenance, repair and cleaning, the use of an FOB can cost as much as $312 per procedure. The VivaSight DLT (VS-DLT) incorporates a built-in camera, which is aimed at reducing FOB use and its associated costs. In this study, we compared the rate of FOB use when intubating using either a VS-DLT or a conventional DLT (c-DLT).Methods
This is a randomized controlled comparative study performed at a public county teaching hospital. A total of 50 patients were enrolled and randomly assigned to either a c-DLT (n=25) or a VS-DLT (n=25). Th