ABSTRACT: Purpose
To develop and evaluate a fast patient localization tool using megavoltage (MV)-topogram on helical tomotherapy.Methods and materials
Eighty-one MV-topogram pairs for 18 pelvis patients undergoing radiation were acquired weekly under an institutional review board-approved clinical trial. The MV-topogram imaging protocol requires 2 orthogonal acquisitions at static gantry angles of 0 degrees and 90 degrees for a programed scan length. A MATLAB based in-house software was developed to reconstruct the MV-topograms offline. Reference images (digitally reconstructed topograms, digitally reconstructed topograms) were generated using the planning computed tomography and tomotherapy geometry. The MV-topogram based alignment was determined by registering the MV-topograms to