ABSTRACT: Purpose
To report a 3D multi-echo thin-slab stack-of-stars (tsSOS) quiescent-interval slice-selective (QISS) strategy for high-resolution magnetic resonance angiography (MRA) of the entire neck in under seven minutes.Methods
The neck arteries of eight subjects were imaged at 3 Tesla. Multi-echo 3D tsSOS QISS using a FLASH readout was compared with 3D tsSOS FLASH, 2D QISS, 2D TOF, and 3D TOF. A root-mean-square (RMS) combination of echo time images was tested. Evaluation metrics included arterial signal-to-noise ratio (SNR), arterial-to-muscle contrast-to-noise ratio (CNR), and image quality.Results
3D multi-echo tsSOS QISS using a RMS combination of echo time images increased SNR and CNR by 60% and 63% with respect to the reconstruction obtained with the shortest ec