Estimation of CARE Dose 4D quality reference mAs conversion factors for child to adult reference patient in child protocols on Siemens Symbia SPECT-CT systems.


ABSTRACT:

Objectives

CARE Dose 4D modulates mAs through several mechanisms according to patient size and shape, whilst maintaining user-defined reference image quality on Siemens Symbia single-photon emission computed tomography (SPECT)-computed tomography (CT) systems. A 20 kg child reference was used in child protocols prior to software version VB10 and a 75 kg adult thereafter. Quality reference mAs conversion factors are estimated for delivering equivalent mAs to children between two comparable SPECT-CT systems using adult and child references for topogram-based patient-size-related dose level adaptations.

Methods

A child phantom was scanned using child protocols on a Siemens Symbia T16 (child reference) and a Siemens Symbia Intevo Bold (adult reference). On each system, scans of