Quantitative accuracy of virtual non-contrast images derived from spectral detector computed tomography: an abdominal phantom study.
Ontology highlight
ABSTRACT: Dual-energy CT allows for the reconstruction of virtual non-contrast (VNC) images. VNC images have the potential to replace true non-contrast scans in various clinical applications. This study investigated the quantitative accuracy of VNC attenuation images considering different parameters for acquisition and reconstruction. An abdomen phantom with 7 different tissue types (different combinations of 3 base materials and 5 iodine concentrations) was scanned using a spectral detector CT (SDCT). Different phantom sizes (S, M, L), volume computed tomography dose indices (CTDIvol 10, 15, 20 mGy), kernel settings (soft, standard, sharp), and denoising levels (low, medium, high) were tested. Conventional and VNC images were reconstructed and analyzed based on regions of interest (ROI). Mean and s
SUBMITTER: Holz JA
PROVIDER: S-EPMC7725817 | biostudies-literature | 2020 Dec
REPOSITORIES: biostudies-literature