Unknown

Dataset Information

Comparison of internal target volumes defined by three-dimensional, four-dimensional, and cone-beam computed tomography images of a motion phantom.


ABSTRACT:

Background

To explore the variations of the gross tumor volume (GTV) from different three-dimensional computed tomography (3DCT) scan modes and the consistency of internal target volume (ITV) between different 3DCT, cone-beam computed tomography (CBCT) and four-dimensional computed tomography (4DCT) scan, a study using a motion phantom simulating sinusoidal movement was conducted.

Methods

For three 3DCT scan modes: the GTV was contoured, and ITVI was generated on the basis of GTV with a 10-mm margin while ITVII and ITVIII with a 0-mm margin on the motion direction. ITVCBCT and ITVMIP were contoured on the images of CBCT and maximum-intensity projection (MIP) reconstructed 4DCT images. The centroid position shifts of ITVs

SUBMITTER: Wang Y 

PROVIDER: S-EPMC7729323 | biostudies-literature | 2020 Nov

REPOSITORIES: biostudies-literature