The effective coupling coefficient for a completed PIN-PMN-PT array.
Ontology highlight
ABSTRACT: The computation of the electromechanical coupling coefficient (EMCC) of a fully assembled medical ultrasound transducer array is directly computed with closed form expressions. The Levenberg-Marquardt non-linear regression algorithm (LMA) is employed to help confirm the EMCC calculated prediction (kEFF) and provide statistical insights. The complex electrical impedance spectra of a 1-3 composite array with two matching layers operating at a 3.75 MHz center frequency using PIN-PMN-PT single crystal material is measured in air both before and after oven heating at 160 °C for 15 min. The oven heating produces changes in the EMCC of -4.9%, clamped dielectric constant of -11%, and effective transducer longitudinal velocity of -2.5%. Utilizing the pre- and post-heating array impedance