Ontology highlight
ABSTRACT: Purpose
To investigate the feasibility and practicality of ultraviolet (UV) germicidal irradiation of the inner bore of a computed tomography (CT) gantry as a means of viral decontamination.Method
A UV lamp (PADNUT 38 W, 253 nm UV-C light tube) and UV-C dosimeter (GENERAL UV-C Digital Light Meter No. UV512C) were used to measure irradiance throughout the inner bore of a CT scanner gantry. Irradiance (units μW/cm2 ) was related to the time required to achieve 6-log viral kill (10-6 survival fraction).Results
A warm-up time of ~120 s was required for the lamp to reach stable irradiance. Irradiance at the scan plane (z = 0 cm) of the CT scanner was 580.9 μW/cm2 , reducing to ~350 μW/cm2 at z = ±20 cm toward the front or back