An evaluation of the effect of tube potential on clinical image quality using direct digital detectors for pelvis and lumbar spine radiographs.


ABSTRACT:

Introduction

High kVp techniques, 15% or 10-kVp rules, are well-known dose reduction methods. Traditionally, the use of high tube potential (i.e. increased kVp) is associated with decreased radiographic contrast and overall image quality. Recent studies suggest contrast and image quality are not heavily reliant on kVp with digital systems. This study aims to assess the effects of the high tube potential technique on clinical radiographic image quality when using digital systems, to validate high kVp as a dose saving technique.

Methods

A selection of comparable pelvis and lumbar spine radiographs were collected from the hospital's picture archiving and communication system (PACS), with technical factors recorded. All clinical radiographs were assessed by 5 senior radiographer