ABSTRACT: Purpose
The aim of this study was to provide a comprehensive assessment of patient intrafraction motion in linac-based frameless stereotactic radiosurgery (SRS) and radiotherapy (SRT).Methods
A retrospective review was performed on 101 intracranial SRS/SRT patients immobilized with the Klarity stereotactic thermoplastic mask (compatible with the Brainlab frameless stereotactic system) and aligned on a 6 Degree of Freedom (DoF) couch with the Brainlab ExacTrac image guidance system. Both pretreatment and intrafraction correction data are provided as observed by the ExacTrac system. The effects of couch angle and treatment duration on positioning outcomes are also explored.Results
Initial setup data for patients is shown to vary by up to ±4.18 mm, ±2.97°, but when cor