Feasibility of non-contact cardiorespiratory monitoring using impulse-radio ultra-wideband radar in the neonatal intensive care unit.


ABSTRACT:

Background

Current cardiorespiratory monitoring equipment can cause injuries and infections in neonates with fragile skin. Impulse-radio ultra-wideband (IR-UWB) radar was recently demonstrated to be an effective contactless vital sign monitor in adults. The purpose of this study was to assess heart rates (HRs) and respiratory rates (RRs) in the neonatal intensive care unit (NICU) using IR-UWB radar and to evaluate its accuracy and reliability compared to conventional electrocardiography (ECG)/impedance pneumography (IPG).

Methods

The HR and RR were recorded in 34 neonates between 3 and 72 days of age during minimal movement (51 measurements in total) using IR-UWB radar (HRRd, RRRd) and ECG/IPG (HRECG, RRIPG) simultaneously. The radar signals were processed in real time using