Diagnostic accuracy of a bovine specific electronic beta-hydroxybutyrate handheld meter in fresh blood and stored serum samples.
Ontology highlight
ABSTRACT: This study aims to evaluate the diagnostic accuracy of the bovine specific beta-hydroxybutyrate (BHB) meter Nova Vet (NVET). We evaluated the accuracy and agreement of the NVET in fresh blood and thawed serum with the reference laboratory assay; and the repeatability, the interference by anticoagulants, and the optimum slope calibration factor. Individual blood samples were collected from 200 Holstein and crossbred cows between 3 and 14 days post-calving from 13 dairy herds in Minnesota. Using a laboratory assay with a cut point of 1.2 BHB mmol/L hyperketonemia prevalence was 10.6% (95% CI: 6.7, 15.8). The sensitivity of NVET in blood and serum was 100.0% while the specificity was 98.3 and 97.7% respectively. The agreement between NVET and the laboratory assay was the highest
SUBMITTER: Rodriguez Z
PROVIDER: S-EPMC7807146 | biostudies-literature | 2021 Mar