Multi-energy spectral photon-counting computed tomography (MARS) for detection of arthroplasty implant failure.
Ontology highlight
ABSTRACT: To determine whether state-of-the-art multi-energy spectral photon-counting computed tomography (MARS) can detect knee arthroplasty implant failure not detected by standard pre-operative imaging techniques. A total knee arthroplasty (TKA) removed from a patient was reviewed. The extracted prosthesis [NexGen Legacy Posterior Stabilized (LPS) TKA] was analyzed as were pre-operative imaging examination and compared with a MARS-CT examination obtained of the extracted TKA prosthesis. Radiographs, fluoroscopy, ultrasound and MRI preoperatively did not reveal the cause of the implant failure. MARS CT images of the extracted prosthesis clearly showed the presence of posteromedial polyethylene and tibial tray wear which is compatible with the clinical appearance of the extracted TKA. MARS can iden