ABSTRACT: Background
Bovine pericardium can be used for cardiovascular repair surgeries, but challenges involving biocompatibility and durability remain. This study aimed to carry out pre-clinical testing of aortic valve replacement using an aortic valve prosthesis made of bovine pericardium modified with glutaraldehyde (GA) and 2,3-butanediol (BD).Methods
The mechanical, plasma protein adsorption, platelet adhesion, collagenase digestion, and ninhydrin properties of the material (control vs. GA vs. GA + BD) were tested. All 3 tissues were implanted in rats and observed after 8 weeks under microscopy with alizarin red staining for calcification. Aortic valves made from the fully-treated material were implanted in sheep. A commercial bioprosthesis was used as control. Ef
SUBMITTER: Ren K
PROVIDER: S-EPMC7812161 | biostudies-literature | 2020 Dec