Unknown

Dataset Information

Portable smartphone-integrated paper sensors for fluorescence detection of As(III) in groundwater.


ABSTRACT: Arsenic contamination in groundwater is a supreme environmental problem, and levels of this toxic metalloid must be strictly monitored by a portable, sensitive and selective analytical device. Herein, a new system of smartphone-integrated paper sensors with Cu nanoclusters was established for the effective detection of As(III) in groundwater. For the integration system, the fluorescence emissive peak of Cu nanoclusters at 600 nm decreased gradually with increasing As(III) addition. Meanwhile, the fluorescence colour also changed from orange to colourless, and the detection limit was determined as 2.93 nM (0.22 ppb) in a wide detection range. The interfering ions also cannot influence the detection selectivity of As(III). Furthermore, the portable paper sensors based on Cu nanoclusters were

SUBMITTER: Liu S 

PROVIDER: S-EPMC7813225 | biostudies-literature | 2020 Dec

REPOSITORIES: biostudies-literature