Performance of two rapid point of care SARS-COV-2 antibody assays against laboratory-based automated chemiluminescent immunoassays for SARS-COV-2 IG-G, IG-M and total antibodies.


ABSTRACT:

Introduction

We evaluated two SARS-CoV-2 antibody point-of-care tests (POCTs) (Abbott Panbio COVID-19 IgG/IgM and Roche SARS-CoV-2 Rapid Antibody tests) and compared the results to their respective chemiluminescent immunoassays (CLIAs) (Abbott Architect IgM, Architect IgG, Roche Cobas total antibody assays).

Method

200 pre-pandemic sera and 48 samples positive for various conditions (18 viral hepatitis, 18 dengue, 11 ANA and 1 dsDNA) were used as controls and to assess for cross-reactivity. Anonymised residual leftover sera positive for SARS-CoV-2 on RT-PCR were recruited as cases (n ​= ​133). The sensitivity/specificity/cross-reactivity/positive predictive value (PPV)/negative predictive value (NPV) of the POCTs were assessed. Concordance between the POCTs and chemiluminesc

SUBMITTER: Lau CS 

PROVIDER: S-EPMC7816621 | biostudies-literature | 2021 Mar