Evaluation of the analytical performance of endocrine analytes using sigma metrics.


ABSTRACT:

Background

(a) To evaluate the clinical performance of endocrine analytes using the sigma metrics (σ) model. (b) To redesign quality control strategies for performance improvement.

Methods

The sigma values of the analytes were initially evaluated based on the allowable total error (TEa), bias, and coefficient of variation (CV) at QC materials level 1 and 2 in March 2018. And then, the normalized QC performance decision charts, personalized QC rules, quality goal index (QGI) analysis, and root causes analysis (RCA) were performed based on the sigma values of the analytes. Finally, the sigma values were re-evaluated in September 2018 after a series of targeted corrective actions.

Results

Based on the initial sigma values, two analytes (FT3 and TSH) with σ > 6, only need