ABSTRACT: Background
Incubators and radiant warmers are essential equipment in neonatal care, but the typical 1,500 to 35,000 USD cost per device makes it unaffordable for many units in low and middle-income countries. We aimed to determine whether stable preterm infants could maintain thermoregulation for 48 h in a low-cost incubator (LCI).Methods
The LCI was constructed using a servo-heater costing 200 USD and cardboard infant-chamber. We conducted this open-labeled non-inferiority randomized controlled trial in a tertiary level teaching hospital in India from May 2017 to March 2018. Preterm infants on full feeds and receiving incubator or radiant warmer care were enrolled at 32 to 36 weeks post-menstrual age. We enrolled 96 infants in two strata (Strata-1< 33 weeks, Strata-2 ≥ 33