ABSTRACT: Objectives
We aimed to evaluate the accuracy of the Panbio™ Ag-RDT at primary health care (PHC) centers and test sites in symptomatic patients and close contacts, using the Reverse-Transcription Polymerase Chain Reaction (RT-PCR) test as the gold standard.Methods
The study was conducted in four PHC centers and two test sites in Mallorca, Spain. Consecutive patients older than 18 years, attending the sites for RT-PCR testing were included. Two nasopharyngeal samples were collected, one for RT-PCR and the other was processed on-site using the Panbio™ rapid antigen test kit for SARS-CoV-2. The sensitivity and specificity were calculated using RT-PCR as the reference, and the predictive values using the pretest probability results for each analyzed group.Findings