Quantitative imaging of bone remodeling in patients with a unicompartmental joint unloading knee implant (ATLAS Knee System)-effect of metal artifacts on a SPECT-CT-based quantification.


ABSTRACT:

Background

SPECT-CT using radiolabeled phosphonates is considered a standard for assessing bone metabolism (e.g., in patients with osteoarthritis of knee joints). However, SPECT can be influenced by metal artifacts in CT caused by endoprostheses affecting attenuation correction. The current study examined the effects of metal artifacts in CT of a specific endoprosthesis design on quantitative hybrid SPECT-CT imaging. The implant was positioned inside a phantom homogenously filled with activity (955 MBq 99mTc). CT imaging was performed for different X-ray tube currents (I = 10, 40, 125 mA) and table pitches (p = 0.562 and 1.375). X-ray tube voltage (U = 120 kVp) and primary collimation (16 × 0.625 mm) were kept constant for all scans. The CT reconstruction was performed w