Ratchet mechanism selectively causing idiopathic macrodislodgement of an active-fixation coronary sinus lead: a case report.


ABSTRACT:

Background

'Idiopathic' lead macrodislodgement may be due to Twiddler's syndrome depending on active twisting of pulse generator within subcutaneous pocket. All leads are involved, at any time from implantation, and frequently damaged. In the past few years, a reel syndrome was also observed: retraction of pacemaker leads into pocket without patient manipulation, owing to lead circling the generator. In other cases, a 'ratchet' mechanism has been postulated. Reel and ratchet mechanisms require loose anchoring, occur generally briefly after implantation, with non-damaged leads. We report the first case of an active-fixation coronary sinus lead selective macrodislodgement involving such ratchet mechanism.

Case summary

A 65-year-old man underwent biventricular defibrillator dev

SUBMITTER: Golzio PG 

PROVIDER: S-EPMC7891260 | biostudies-literature | 2020 Dec