Ontology highlight
ABSTRACT: Introduction
While wideband segmented, breath-hold late gadolinium-enhancement (LGE) cardiovascular magnetic resonance (CMR) has been shown to suppress image artifacts associated with cardiac-implanted electronic devices (CIEDs), it may produce image artifacts in patients with arrhythmia and/or dyspnea. Single-shot LGE is capable of suppressing said artifacts. We sought to compare the performance of wideband single-shot free-breathing LGE against the standard and wideband-segmented LGEs in CIED patients.Methods and results
We retrospectively identified all 54 consecutive patients (mean age: 61 ± 15 years; 31% females) with CIED who had undergone CMR with standard segmented, wideband segmented, and/or wideband single-shot LGE sequences as part of quality assurance for determ
SUBMITTER: Schwartz SM
PROVIDER: S-EPMC7902469 | biostudies-literature | 2021 Jan
REPOSITORIES: biostudies-literature