The Influence of Skin Thickness on Flash Glucose Monitoring System Accuracy in Dogs with Diabetes Mellitus.


ABSTRACT: A flash glucose monitoring system (FGMS) has been validated for use in diabetic dogs. However, it is unknown whether skin thickness affects FGMS measurements. The aim of this study was to evaluate whether FGMS accuracy is affected by skin thickness. Fourteen client-owned diabetic dogs on insulin treatment were prospectively enrolled in the study. The dogs were divided into two groups according to their ultrasound-measured skin thickness: dogs with skin thickness < 5 mm (Group 1) and dogs with skin thickness > 5 mm (Group 2). On days 1, 7 and 14, glucose curves were obtained simultaneously using the FGMS and a validated portable blood glucose meter. Paired measurements were used to calculate the mean bias and to determine accuracy according to ISO 15197:2013 criteria. The mean bias was sign