Correction of lateral response artifacts from flatbed scanners for dual-channel radiochromic film dosimetry.
Ontology highlight
ABSTRACT: In this study, we evaluated the inter-unit variability of the lateral response artifact for multiple flatbed scanners, focusing on the dual-channel method, and investigated the correction method of the lateral non-uniformity. Four scanners with A3+ paper-size and five scanners with A4 paper-size were evaluated. To generate the dose-response curves, small pieces of the Gafchromic EBT3 and EBT-XD films were irradiated, and five of the pieces were repeatedly scanned by moving them on the scanner to evaluate the lateral non-uniformity. To calculate the dose distribution accounting for the lateral non-uniformity, linear functions of the correction factor, representing the difference between the pixel values at offset position and the scanner midline, were calculated for red and blue color chann
SUBMITTER: Akino Y
PROVIDER: S-EPMC7948896 | biostudies-literature | 2021 Mar