Volumetric imaging of myelin in vivo using 3D inversion recovery-prepared ultrashort echo time cones magnetic resonance imaging.


ABSTRACT: Direct myelin imaging is promising for characterization of multiple sclerosis (MS) brains at diagnosis and in response to therapy. In this study, a 3D inversion recovery-prepared ultrashort echo time cones (IR-UTE-Cones) sequence was used for both morphological and quantitative imaging of myelin on a clinical 3 T scanner. Myelin powder phantoms with different myelin concentrations were imaged with the 3D UTE-Cones sequence and it showed a strong correlation between concentrations and UTE-Cones signals, demonstrating the ability of the UTE-Cones sequence to directly image myelin in the brain. Quantitative myelin imaging with multi-echo IR-UTE-Cones sequences show similar T2 * values for a D2 O-exchanged myelin phantom (T2 * = 0.33 ± 0.04 ms), ex vivo brain s