Validity of the Apple Watch® for monitoring push counts in people using manual wheelchairs.


ABSTRACT: Objective: A recent Apple Watch® activity-monitoring innovation permits manual wheelchair users to monitor daily push counts. This study evaluated the validity of the Apple Watch® push count estimate.Design: Criterion validity.Setting: Southern Finland and Southeast Queensland, Australia.Participants: Twenty-six manual wheelchair users from Finland and Australia were filmed completing a standardized battery of activities while wearing the Apple Watch® (dominant wrist).Outcome Measures: Wheelchair pushes as determined by the Apple Watch® were compared to directly observed pushes.Results: Agreement between Apple Watch® push counts and directly observed pushes was evaluated using Intraclass correlation coefficients (ICC), Pearson correlations and Bland-Altman analyses. Apple Watch® pushes and directly observed push counts were strongly correlated (ICC = 0.77, P < 0.01) (r = 0.84, P < 0.01). Bland Altman plots indicated that the Apple Watch® underestimated push counts (M = -103; 95% ULoA = 217; LLoA = -423 pushes). Mean absolute percentage error was 13.5% which is comparable to studies evaluating agreement between pedometer-based step counts and directly observed steps.Conclusion: Apple Watch® push-count estimates are acceptable for personal, self-monitoring purposes and for research entailing group-level analyses, but less acceptable where accurate push-count measures for an individual is required.

SUBMITTER: Karinharju KS 

PROVIDER: S-EPMC7952070 | biostudies-literature | 2021 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Validity of the Apple Watch<sup>®</sup> for monitoring push counts in people using manual wheelchairs.

Karinharju Kati S KS   Boughey Alexandra M AM   Tweedy Sean M SM   Clanchy Kelly M KM   Trost Stewart G SG   Gomersall Sjaan R SR  

The journal of spinal cord medicine 20190227 2


<b>Objective:</b> A recent Apple Watch<sup>®</sup> activity-monitoring innovation permits manual wheelchair users to monitor daily push counts. This study evaluated the validity of the Apple Watch<sup>®</sup> push count estimate.<b>Design:</b> Criterion validity.<b>Setting</b>: Southern Finland and Southeast Queensland, Australia.<b>Participants</b>: Twenty-six manual wheelchair users from Finland and Australia were filmed completing a standardized battery of activities while wearing the Apple W  ...[more]

Similar Datasets

| S-EPMC7886944 | biostudies-literature
| S-EPMC10383699 | biostudies-literature
| S-EPMC5489335 | biostudies-literature
| S-EPMC11585612 | biostudies-literature
| S-EPMC11306942 | biostudies-literature
| S-EPMC9780663 | biostudies-literature
| S-EPMC10605392 | biostudies-literature
| S-EPMC6689219 | biostudies-literature
| S-EPMC10368659 | biostudies-literature
| S-EPMC8144193 | biostudies-literature