Comparison of Autof ms1000 and Bruker Biotyper MALDI-TOF MS Platforms for Routine Identification of Clinical Microorganisms.


ABSTRACT: Matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) is widely used in clinical microbiology laboratories because it is cost-effective, reliable, and fast. This study is aimed at comparing the identification performance of the recently developed Autof ms1000 (Autobio, China) with that of the Bruker Biotyper (Bruker Daltonics, Germany). From January to June 2020, 205 preserved strains and 302 clinical isolates were used for comparison. Bacteria were tested with duplicates of the direct transfer method, and formic acid extraction was performed if the results were not at the species level. Fungi were tested with formic acid extraction followed by ethanol extraction methods. 16S rRNA or ITS region sequence analysis was performed on isolates that could not