Diagnosis of sleep apnea without sensors on the patient's face.


ABSTRACT:

Study objectives

Thermistors, nasal cannulas, and respiratory inductance plethysmography (RIP) are the recommended reference sensors of the American Academy of Sleep Medicine (AASM) for the detection and characterization of apneas and hypopneas; however, these sensors are not well tolerated by patients and have poor scorability. We evaluated the performance of an alternative method using a combination of tracheal sounds (TSs) and RIP signals.

Methods

Consecutive recordings of 70 adult patients from the Pays de la Loire Sleep Cohort were manually scored in random order using the AASM standard signals and the combination TS and RIP signals, without respiratory sensors placed on the patient's face. The TS-RIP scoring used the TS and RIP-flow signals for detection of apneas and