ABSTRACT: Background and purpose
Aneurysm clips need to be tested at 3T to characterize MR imaging concerns, including magnetic field interactions, MR imaging-related heating, and artifacts. Therefore, we evaluated these risks for aneurysm clips.Materials and methods
Three different MP35N aneurysm clips (Codman Slim-Line Aneurysm Clip, straight, blade length 25-mm; Codman Slim-Line Aneurysm Clip Graft, 5-mm diameter x 5-mm width; Codman Slim-Line Aneurysm Clip, reinforcing 30 degrees angle, 6-mm x 18-mm) that represented the largest mass for 155 additional clips made from MP35N were tested. The clips were evaluated at 3T for magnetic field interactions, heating, and artifacts. We studied MR imaging-related heating, placing the clip in a gelled-saline-filled phantom with MR imaging pe