Use of Standardized Uptake Value Ratios Decreases Interreader Variability of [18F] Florbetapir PET Brain Scan Interpretation.


ABSTRACT:

Background and purpose

Fluorine-18 florbetapir is a recently developed β-amyloid plaque positron-emission tomography imaging agent with high sensitivity, specificity, and accuracy in the detection of moderate-to-frequent cerebral cortical β-amyloid plaque. However, the FDA has expressed concerns about the consistency of interpretation of [(18)F] florbetapir PET brain scans. We hypothesized that incorporating automated cerebral-to-whole-cerebellar standardized uptake value ratios into [(18)F] florbetapir PET brain scan interpretation would reduce this interreader variability.

Materials and methods

This randomized, blinded-reader study used previously acquired [(18)F] florbetapir scans from 30 anonymized patients who were enrolled in the Alzheimer's Disease Neuroimaging Initia