Unknown

Dataset Information

Initial Feasibility and Clinical Implementation of Daily MR-Guided Adaptive Head and Neck Cancer Radiation Therapy on a 1.5T MR-Linac System: Prospective R-IDEAL 2a/2b Systematic Clinical Evaluation of Technical Innovation.


ABSTRACT:

Purpose

This prospective study is, to our knowledge, the first report of daily adaptive radiation therapy (ART) for head and neck cancer (HNC) using a 1.5T magnetic resonance imaging-linear accelerator (MR-linac) with particular focus on safety and feasibility and dosimetric results of an online rigid registration-based adapt to position (ATP) workflow.

Methods and materials

Ten patients with HNC received daily ART on a 1.5T/7MV MR-linac, 6 using ATP only and 4 using ATP with 1 offline adapt-to-shape replan. Setup variability with custom immobilization masks was assessed by calculating the mean systematic error (M), standard deviation of the systematic error (Σ), and standard deviation of the random error (σ) of the isocenter shifts. Quality assurance was performed with a cy

SUBMITTER: McDonald BA 

PROVIDER: S-EPMC7965360 | biostudies-literature | 2021 Apr

REPOSITORIES: biostudies-literature