ABSTRACT: Goal
The objective of this paper is to investigate if the use of a B-scan ultrasound imaging system can reduce the energy requirements, and hence the power-dissipation requirements to support wireless bio-telemetry at an implantable device.Methods
B-scan imaging data were acquired using a commercial 256-element linear ultrasound transducer array which was driven by a commercial echoscope. As a transmission medium, we used a water-bath and the operation of the implantable device was emulated using a commercial-off-the-shelf micro-controller board. The telemetry parameters (e.g. transmission rate and transmission power) were wirelessly controlled using a two-way radio-frequency transceiver. B-scan imaging data were post-processed using a maximum-threshold decoder and the qual