Unknown

Dataset Information

A comparison of the analysis of 3 types of body fluids using the XN-350 hematology analyzer versus light microscopy assessment.


ABSTRACT:

Abstract

We evaluated the capacity of the XN-350 instrument to analyze 3 different types of body fluid samples under "body fluid mode."The performance of XN-350 was evaluated in terms of precision, carryover, limit of blank, limit of detection, limit of quantification, and linearity. Cell enumeration and differential data produced by the XN-350 were compared to manual chamber counting results in 63 cerebrospinal fluid (CSF), 51 ascitic fluid, and 51 pleural fluid (PF) samples. Comparisons between XN-350 versus Cytospin data were also performed in PF samples.The precision, carry-over, limit of blank, and linearity of the XN-350 were acceptable. The limits of detection for white blood cells (WBCs) and red blood cells were 1.0/μL, and 1,000.0/μL, respectively; the corresponding limits

SUBMITTER: Lee J 

PROVIDER: S-EPMC7982217 | biostudies-literature | 2021 Mar

REPOSITORIES: biostudies-literature