Performance of three different continuous glucose monitoring systems in children with type 1 diabetes during a diabetes summer camp.


ABSTRACT: The aim of this study was to assess accuracy of the three most commonly used continuous glucose monitoring (CGM) systems in almost real-life situation during a diabetes camp in children with type 1 diabetes (T1D) aged 9-14 years. Data was gathered during a 2-week summer camp under physicians' supervision. Out of 38 participating children with T1D (aged: 11.0 [9.9; 12.1] years; 57% girls, mean HbA1c 7.2 [6.9; 7.7] %,) 37 wore a CGM system (either Abbott FreeStyle Libre (FSL), Dexcom G6 (DEX) or Medtronic Enlite (ENL)) throughout the camp. All concomitantly available data pairs of capillary glucose measurements and sensor values were used for the analysis. Mean absolute relative difference (MARD) was calculated and Parkes Error Grid analyses were done for all three systems used. In total 207