ABSTRACT: Purpose
Multi-transmit MRI systems are typically equipped with dedicated hardware to sample the reflected/lost power in the transmit channels. After extensive calibration, the amplitude and phase of the signal at the feed of each array element can be accurately determined. However, determining the phase is more difficult and monitoring errors can lead to a hazardous peak local specific absorption rate (pSAR10g ) underestimation. For this purpose, methods were published for online maximum potential pSAR10g estimation without relying on phase monitoring, but these methods produce considerable overestimation. We present a trigonometric maximization method to determine the actual worst-case pSAR10g without any overestimation.Theory and method
Th