Continuous 24-hour measurement of intraocular pressure in millimeters of mercury (mmHg) using a novel contact lens sensor: Comparison with pneumatonometry.


ABSTRACT:

Purpose

To address the unmet need of continuous IOP monitoring, a Pressure-Measuring Contact Lens (PMCL) was developed to measure IOP in millimeters of mercury (mmHg) continuously over 24 hours. The present study assessed the reliability of the novel PMCL.

Methods

In this prospective open-label clinical study, healthy and open-angle glaucoma (OAG) subjects were fitted with the PMCL, and pneumatonometry was performed on study eyes (in absence of the PMCL) and on fellow eyes before, during, and after provocative tests. The primary outcome measures were (1) mean IOP difference between same-eye measurements, and (2) percentage of timepoints at which IOP measured by the PMCL was within 5 mmHg of that measured by pneumatonometry in the fellow eye.

Results

Eight subjects wer