Unknown

Dataset Information

0

Improving MiniHip femoral prosthesis positioning using a cross-laser projection system in total hip arthroplasty by an anterolateral supine approach.


ABSTRACT:

Background

The authors developed a cross-laser projection system (CLP) to place a femoral neck-sparing short stem using the minimally invasive anterolateral supine approach in total hip arthroplasty. This study aimed to verify the utility of CLP.

Methods

Thirty joints were assessed with the MiniHip (Corin). The authors compared femoral component implantation with a patient-specific femoral osteotomy guide (PSG) for the femoral neck-cut (PSG group), with the CLP attached to the rasp handle to irradiate the cross-laser to the target of PSG (CLP group), and without PSG or CLP (control group).

Results

In the CLP group, the positional deviation of anteversion, anterior/posterior tilt and varus/valgus placement of the stem postoperatively were 1.8° ± 0.2°, 2.0° ± 2.0° and 2.0° ± 0.1°, respectively. The positional deviation of anteversion (p < 0.001) and anterior/posterior tilt (p = 0.036) were significantly smaller than those in the other groups.

Conclusions

CLP improves the accuracy of MiniHip femoral prosthesis placement.

SUBMITTER: Fujii H 

PROVIDER: S-EPMC7988605 | biostudies-literature | 2021 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

Improving MiniHip femoral prosthesis positioning using a cross-laser projection system in total hip arthroplasty by an anterolateral supine approach.

Fujii Hideki H   Hayama Tetsuo T   Abe Toshiomi T   Takahashi Motoi M   Matsushita Yohei Y   Sato Ryuichi R   Otani Takuya T   Saito Mitsuru M  

The international journal of medical robotics + computer assisted surgery : MRCAS 20210109 2


<h4>Background</h4>The authors developed a cross-laser projection system (CLP) to place a femoral neck-sparing short stem using the minimally invasive anterolateral supine approach in total hip arthroplasty. This study aimed to verify the utility of CLP.<h4>Methods</h4>Thirty joints were assessed with the MiniHip (Corin). The authors compared femoral component implantation with a patient-specific femoral osteotomy guide (PSG) for the femoral neck-cut (PSG group), with the CLP attached to the ras  ...[more]

Similar Datasets

| S-EPMC7364523 | biostudies-literature
| S-EPMC9617043 | biostudies-literature
| S-EPMC4680950 | biostudies-literature
| S-EPMC8891993 | biostudies-literature
| S-EPMC6583161 | biostudies-literature
| S-EPMC2267524 | biostudies-other
| S-EPMC8980438 | biostudies-literature
| S-EPMC6287226 | biostudies-literature
| S-EPMC11776236 | biostudies-literature
| S-EPMC11447336 | biostudies-literature